User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Weston 433 (150 VAC, 25-2400Hz) DATE: 09-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 37 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R F W 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP This meter is tested in the HORIZONTAL position. 1.005 DISP Check meter mechanical zero and adjust if necessary. 1.006 HEAD {* FULL SCALE AND LINEARITY TEST *} 1.007 DISP [32] Connect the 5700A and the UUT as follows: 1.007 DISP [32] 5700A TO 433 1.007 DISP [32] OUTPUT V Ohm A HI .......... UPPER TERMINAL 1.007 DISP [32] OUTPUT V Ohm A LO .......... LOWER TERMINAL 1.008 5700 150.000V 3.000U 400H 2W 2.001 MATH M[1] = MEM1 2.002 5700 130.000V 3.000U 400H 2W 3.001 5700 110.000V 3.000U 400H 2W 4.001 5700 90.000V 3.000U 400H 2W 5.001 5700 70.000V 3.000U 400H 2W 6.001 5700 50.000V 3.000U 400H 2W 7.001 HEAD {* FREQUENCY RESPONSE *} 7.002 MATH MEM = M[1] 7.003 5700 150.000V 3.000U 25H C 2W 8.001 5700 150.000V 3.000U 50H C 2W 9.001 5700 150.000V 3.000U 200H C 2W 10.001 5700 150.000V 3.000U 1kH C 2W 11.001 5700 150.000V 3.000U 2.4kH C 2W 12.001 END